Full text:  

 

Abstract

Today, there is a growing number of patients with implantable active cardiac devices such as pacemakers, cardioverter defibrillators (ICD) and resynchronization therapy devices. Conducting cardiac magnetic resonance imaging (MRI) in such patients remains a problem due to the possible effect of a magnetic field on the device and, accordingly, on the patient. The recent introduction of conditionally safe active implanted heart systems represents a definite step in overcoming the limitations of pre-MRI. However, the described systems require compliance with certain conditions for conducting research and switching to a specialized mode of operation before using in a magnetic field. Our article presents the results of MR studies of patients with various conditionally compatible active implanted heart systems. Studies of patients with implanted pacemakers were carried out in full. But it is worth noting that in the right heart of patients with pacemaker artifacts from implanted electrodes were noted, which did not distort the examination results. Studies of patients with implanted cardioverter-defibrillators were stopped due to pronounced susceptibility artifacts, which made it difficult to evaluate images. Thus, an assessment of the obtained MR images and an analysis of the diagnostic capabilities of cardiac examination for doctors of clinical specialties were carried out.

References

  1. Sergunova K.A., Akhmad E.S., Petryaykin A.V., Kivasev S.A., Semenov D.S., Vasil'ev Yu.A. et al. Magnetic resonance safety for patients with implanted medical devices. The Bulletin of Bakoulev Center for Cardiovascular Diseases. 2019; 20 (4): 313–23 (in Russ.).
  2. Glushko L.A., Bockeria O.L. The devices for electrotherapy of the heart, compatible with magnetic resonance tomography. Annaly Aritmologii (Annals of Arrhythmology). 2015; 12 (4): 225–33 (in Russ.).
  3. Shariya M.A., Ternovoy S.K. MRI-compatible pacemaker (first experience of the examination). Russian Electronic Journal of Radiology. 2013; 3 (2): 92–6 (in Russ.).
  4. Revishvili A.Sh., Neminuschiy N.M., Batalov R.E., Gilyarov M.Yu., Golitsyn S.P., Davtyan K.V. et al. All-russian clinical recommendations for controlling the risk of sudden heart stop and sudden cardiac death, prevention and first aid. Journal of Arrhythmology. 2017; (89): 2–104 (in Russ.).
  5. Berdibekov B.Sh., Aleksandrova S.A., Golukhova E.Z. New evaluation features of ventricular arrhythmia substrate: the role of magnetic resonance imaging. Creative Cardiology. 2019; 13 (3): 217–28 (in Russ.). DOI: 10.24022/1997-3187-2019-13-3-217-228
  6. Ferreira A.M., Mendes L., Soares L., da Graça Correia M., Gil V. Cardiac magnetic resonance in a patient with MRI-conditional pacemaker. Rev. Port. Cardiol. 2013; 32 (2): 159–62. DOI: 10.1016/j.repc.2012.10.006
  7. Koshy A.O., Swoboda P.P.P., Gierula J., Witte K.K. Cardiac magnetic resonance in patients with cardiac resynchronization therapy: is it time to scan with resynchronization on? Europace. 2019; 21 (4): 554–62. DOI: 10.1093/europace/euy299
  8. Edlinger C., Granitz M., Paar V., Jung C., Pfeil A., Eder S. et al. Visualization and appearance of artifacts of leadless pacemaker systems in cardiac MRI: An experimental ex vivo study. Wien Klin. Wochenschr. 2018; 130 (13–14): 427–35. DOI: 10.1007/s00508-018-1334-z
  9. Stevens S.M., Tung R., Rashid S., Gima J., Cote S., Pavez G. et al. Device artifact reduction for magnetic resonanse imaging of patients with implantable cardioverter-defibrillators and ventricular tachycardia: late gadolinium enhancement correlation with electroanatomic mapping. Heart Rhythm. 2014; 11 (2): 289–98. DOI: 10.1016/j.hrthm.2013.10.032
  10. Glikson M., Nielsen J.C., Kronborg M.B., Michowitz Y., Auricchio A., Barbash I.M. et al. 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy. ESC Scientific Document Group. Eur. Heart J. 2021; 42 (35): 3427–520. DOI: 10.1093/eurheartjehab364

About Authors

  • Maria A. Shlyappo, Cand. Med. Sci., Radiologist; ORCID
  • Svetlana A. Aleksandrova, Cand. Med. Sci., Senior Researcher; ORCID
  • Vladimir N. Makarenko, Dr. Med. Sci., Professor, Head of Department; ORCID

Chief Editor

Leo A. Bockeria, MD, PhD, DSc, Professor, Academician of Russian Academy of Sciences, President of Bakoulev National Medical Research Center for Cardiovascular Surgery